Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Goerz SE400: (1 year) CAL VER /5500 DATE: 04-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 24 NUMBER OF LINES: 195 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # ABB GOERZ AG SE 400 Service manual No.331 0073 03 # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP This procedure was intended to check the performance 1.004 DISP charistics of the SE400 with an 884 0811 00 plug-in 1.004 DISP measurement unit, and a time unit. 1.005 DISP Connect UUT to AC power and turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP Set the WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP This procedure will also test the TIME UNIT for 1.007 DISP the GOERZ Instruments model ABB 884. 1.008 DISP Set the REC/<0> switch to <0>. 1.008 DISP Set the V/mV switch to V. 1.008 DISP Set the RANGE knob to 1 (V). 1.008 DISP Set the CAL/VER SWITCH to CAL. 1.008 DISP Set the TIME UNIT to 60cm/min. 1.009 DISP Connect the 5500A and the UUT as follows: 1.009 DISP [32] 5500A NORMAL HI to ABB 884 INPUT HI 1.009 DISP [32] 5500A NORMAL LO to ABB 884 INPUT LO 1.010 HEAD {ZERO ADJUSTABILITY} 1.011 DISP Lower the recorder pens and START the chart feed. 1.012 MESS Turn the ZERO control fully clockwise, then fully 1.012 MESS counterclockwise. Check the trace shifts over the 1.012 MESS entire recording range. 1.012 MESS 1.013 EVAL Does {the ZERO control move trace over entire range}? 2.001 MESS While turning the ZERO control fully clockwise and 2.001 MESS counterclockwise, notice the physical stop points are 2.001 MESS past the edges of the chart grid by approximately 1mm. 2.001 MESS 2.002 EVAL Are {the stop points past ends of chart by .5 - 2mm}? 3.001 MESS 3.002 HEAD {RESPONSE TIME} 3.003 DISP Adjust the ZERO control to 0 (left chart edge) 3.003 DISP Set the REC/<0> switch to REC. 3.004 5500 0V S 2W 3.005 DISP Readjust the ZERO control so trace is on 0% line. 3.006 5500 0.900V S 2W 3.007 DISP [D3999]STOP the chart feed. 3.008 MESS The time between the start (0%) and the end (90%) of the 3.008 MESS movement should be approximately .5 seconds (5mm). 3.008 MESS 3.008 MESS The trace should NOT overshoot the 90% point by 3.008 MESS more than 2.5mm. 3.008 MESS 3.009 EVAL Are {the response time & overshoot characteristics correct}? 4.001 HEAD {DC VOLTAGE} 4.002 DISP Set the TIME UNIT to 12cm/min. 4.003 DISP Set the START the chart feed. 4.004 MESS Slew the 5500A for full scale deflection. 4.004 MESS 4.005 HEAD DC VOLTAGE: {1mV Range} 4.006 DISP Set the V/mV to: mV. 4.006 DISP Set the RANGE to 1 (mV). 4.007 5500 1 1.000mV 0.010U 2W 5.001 HEAD DC VOLTAGE: {2mV Range} 5.002 DISP Set the RANGE to 2 (mV). 5.003 5500 2 2.000mV 0.010U 2W 6.001 HEAD DC VOLTAGE: {5mV Range} 6.002 DISP Set the RANGE to 5 (mV). 6.003 5500 5 5.000mV 0.5% 2W 7.001 HEAD DC VOLTAGE: {10mV Range} 7.002 DISP Set the RANGE to 10 (mV). 7.003 5500 10 10.00mV 0.5% 2W 8.001 HEAD DC VOLTAGE: {20mV Range} 8.002 DISP Set the RANGE to 20 (mV). 8.003 5500 20 20.00mV 0.5% 2W 9.001 HEAD DC VOLTAGE: {50mV Range} 9.002 DISP Set the RANGE to 50 (mV). 9.003 5500 50 50.00mV 0.5% 2W 10.001 HEAD DC VOLTAGE: {100mV Range} 10.002 DISP Set the RANGE to 100 (mV). 10.003 5500 100 100.0mV 0.5% 2W 11.001 HEAD DC VOLTAGE: {200mV Range} 11.002 DISP Set the RANGE to 200 (mV). 11.003 5500 200 200.0mV 0.5% 2W 12.001 HEAD DC VOLTAGE: {500mV Range} 12.002 DISP Set the RANGE to 500 (mV). 12.003 5500 500 500.0mV 0.5% 2W 13.001 HEAD DC VOLTAGE: {1V Range} 13.002 DISP Set the V/mV to V. 13.002 DISP Set the RANGE to 1 (V). 13.003 5500 1 1.000V 0.5% 2W 14.001 HEAD DC VOLTAGE: {2V Range} 14.002 DISP Set the RANGE to 2 (V). 14.003 5500 2 2.000V 0.5% 2W 15.001 HEAD DC VOLTAGE: {5V Range} 15.002 DISP Set the RANGE to 5 (V). 15.003 5500 5 5.000V 0.5% 2W 16.001 HEAD DC VOLTAGE: {10V Range} 16.002 DISP Set the RANGE to 10 (V). 16.003 5500 10 10.00V 0.5% 2W 17.001 HEAD DC VOLTAGE: {20V Range} 17.002 DISP Set the RANGE to 20 (V). 17.003 5500 20 20.00V 0.5% 2W 18.001 HEAD DC VOLTAGE: {50V Range} 18.002 DISP Set the RANGE to 50 (V). 18.003 5500 50 50.00V 0.5% 2W 19.001 HEAD DC VOLTAGE: {100V Range} 19.002 DISP Set the RANGE to 100 (V). 19.003 5500 100 100.0V 0.5% 2W 20.001 HEAD DC VOLTAGE: {200V Range} 20.002 DISP Set the RANGE to 200 (V). 20.003 5500 200 200.0V 0.5% 2W 21.001 MESS 21.002 DISP STOP the chart feed 21.003 HEAD {PEN RECORDING} 21.004 DISP Set the RANGE knob to 1 (V). 21.004 DISP Set the CAL/VER SWITCH to CAL. 21.004 DISP Set the TIME UNIT to 30cm/min. 21.005 DISP START the chart feed. 21.006 5500 0V S 2W 21.007 DISP Adjust the ZERO control to center the trace (50%). 21.008 5500 -0.5V S 2W 21.009 PORT [D1999] 21.010 5500 0V S 2W 21.011 PORT [D3999] 21.012 5500 0.5V S 2W 21.013 PORT [D1999] 21.014 5500 0V S 2W 21.015 DISP STOP the chart feed. 21.016 MESS Check the preceeding trace returns to the center after 21.016 MESS going to full scale in both directions. 21.016 MESS 21.016 MESS The deadband is the offset from center when returned to 21.016 MESS center. 21.016 MESS 21.017 EVAL Is {the DEADBAND (at chart center) <.3mm}? 22.001 HEAD {LINEARITY} 22.002 DISP Set the TIME UNIT to 60cm/min 22.002 DISP 22.002 DISP START the chart feed. 22.003 5500 0.9Vpp 0.1H TI S 2W 22.004 DISP Wait for several cycles of the triangle waveform to be 22.004 DISP recorded, then STOP the chart feed. 22.005 5500 * S 22.006 MESS Check the recorded waveform is straight lines and has no 22.006 MESS irregularities. 22.006 MESS 22.007 EVAL Is {the recording LINEAR +/-.1cm}? 23.001 5500 * S 23.002 HEAD {FEEDRATE} 23.003 DISP START the chart feed. 23.003 DISP 23.004 5500 0.9Vpp 0.01H SQ S 2W 23.005 DISP Let the chart run for ONE COMPLETE CYCLE of the 23.005 DISP squarewave, then STOP the chart feed (this will take 23.005 DISP approx. 100 seconds). 23.006 5500 * S 23.007 MESS Measure the number of centimeters it took the 23.007 MESS recording to complete ONE cycle. 23.007 MESS 23.008 EVAL Is {the CHART FEED 100cm +/-.01cm}? 24.001 END